Dräger Babylog 8000 plus Instructions For Use Manual page 88

Intensive care ventilator for neonates
Hide thumbs Also See for Babylog 8000 plus:

Advertisement

Trends
Displaying lung-mechanics parameters
The patient's resistance and compliance are
calculated using the linear-regression method.
Information on calculations can be found on
page 167.
After a manual inspiration, the display of
waveforms is automatically stopped for one
minute. The waveforms and associated measured
values can be evaluated together.
1 On the main screen, press the Meas key.
2 Press the RC key.
C
Dynamic compliance of the respiratory
system
R
Airway resistance including the tube
C
/C
Index indicating if the lungs are being
20
overinflated. If C
may be overinflated. Can only be used
with ventilation modes without plateau.
Tc
Time constant of the respiratory sys-
tem in milliseconds
r
Correlation coefficient of linear regres-
sion
If the warning symbol (A) is displayed
next to the value for r, the measured
values may be distorted, e.g., due to a
leakage.
88
A
/C <0.8 the lungs
20
Displaying measured volume values
1 On the main screen, press the Meas key.
2 Press the Vol key.
The following measured volume values are
displayed (example):
MV
Expiratory minute volume
spont
Percentage of spontaneous breathing
to minute volume
Leak
Tube leakage
VT
Expiratory tidal volume of the previous
respiratory cycle
Displaying combinations of measured values
1 On the main screen, press the Meas key.
2 Press the MV O
P key.
2
The following measured values are displayed
(example):
MV
Expiratory minute volume
FiO
Measured inspiratory O
2
Mean
Mean airway pressure of previous
respiratory cycle
Instructions for use Babylog 8000 plus SW 5.n
concentration
2