Technical data
Insp. Resistance
Exp. Resistance
Breathing gas humidifier
Resistance
Insp. Resistance
Exp. Resistance
Additional functions
Emergency breathing valve
Safety valve
Subatmospheric pressure in HFV ventilation mode Airway pressure may be subatmospheric during
Displayed measured values
Airway pressure measurement
Sensor
Range
Zero point error
Reinforcement error
Measurement conditions
Peak Inspiratory Pressure
Positive end-expiratory pressure
Mean pressure
Range
Resolution
Airway pressure
Bar display (on device)
Range
Resolution
134
<11 mbar (<11 cmH
O) at 30 L/min
2
<5 mbar (<5 cmH
O) at 30 L/min
2
In accordance with EN ISO 8185
<20 mbar/L/s (<20 cmH
12 mbar/L/s (12 cmH
8 mbar/L/s (8 cmH
2
Opens the breathing system in the event of a fault
Opens at 90 to 110 mbar (90 to 110 cmH
expiration, however it will not fall below –4 mbar
(–4 cmH
O) in the event of a fault.
2
2 piezo-resistive pressure sensors
–10 to 100 mbar (–10 to 100 cmH
±1 mbar (±1 cmH
O)
2
±3 % of measured value
In accordance with ISO 80601-2-12
Pinsp
PEEP
Mean
0 to 99 mbar (0 to 99 cmH
0 to 9.9 mbar (0 to 9.9 cmH
(0.1 cmH
O)
2
10 to 99 mbar (10 to 99 cmH
(1 cmH
O)
2
–10 to 80 mbar (–10 to 80 cmH
2 mbar (2 cmH
O)
2
Instructions for use Babylog 8000 plus SW 5.n
O/L/s)
2
O/L/s)
2
O/L/s)
O)
2
O)
2
O)
2
O): 0.1 mbar
2
O): 1 mbar
2
O)
2