WARNING
Risk of fire
The flow sensor can ignite medications or
other substances based on highly flammable
substances.
– Do not nebulize medications or other
substances that are easily flammable or
spray them into the device.
– Do not use substances containing alcohol.
– Do not allow flammable or explosive
substances to enter the breathing system
or the breathing circuit.
WARNING
Risk due to failure of flow measurement
Deposits that were not removed during
reprocessing can damage the measuring
wires in the flow sensor or cause a fire.
– Before inserting the flow sensor check for
visible damage, soiling, and particles.
Repeat this check regularly.
– Replace flow sensors when damaged,
soiled, or not particle-free.
WARNING
Risk of fire
The use of unapproved O
can lead to excess pressure, which can cause
a fire.
– When supplying the ventilator with
oxygen from a compressed gas cylinder,
only use pressure reducers compliant
with ISO 10524.
– Slowly open the pressure reducer
manually. Do not use tools.
Instructions for use Babylog 8000 plus SW 5.n
pressure reducers
2
For your safety and that of your patients
WARNING
Risk of fire
The medical device may catch fire as a result
of oxygen enrichment in the ambient air.
Medical device malfunctions can increase the
O
concentration in the ambient air.
2
The medical device is only suitable for use in
rooms with sufficient ventilation.
WARNING
Risk of fire
Do no use the medical device together with
flammable gases or flammable solutions that
may become mixed with air, oxygen or
nitrous oxide, or other ignition sources, as
the medical device may catch fire.
Do not allow the medical device to come into
contact with sources of ignition.
WARNING
Risk of electric shock
There is a risk of electric shock if the
connectors of the interfaces and the patient
are touched simultaneously.
Do not simultaneously touch the connectors
of the interfaces and the patient.
WARNING
Risk of patient injury
Penetrating liquid may cause malfunction of
the device, which may endanger the patient.
– Do not place any containers with liquid on
or above the device.
– Make sure that no liquid penetrates into
the device when disinfecting surfaces.
13