Dräger Babylog 8000 plus Instructions For Use Manual page 93

Intensive care ventilator for neonates
Hide thumbs Also See for Babylog 8000 plus:

Advertisement

Information on monitoring
Sensors and calibration intervals
The device uses the following sensors for
measurement and monitoring:
Sensor
Calibration
Pressure sen-
Automatic calibration
sor
O
sensor
Automatic calibration every
2
24 hours
Manual calibration after replace-
ment
Flow sensor
Manual calibration:
– After switching on
– After reprocessing
– After replacement
CAUTION
Device function impaired
If the sensors are not calibrated, device function
may be impaired.
Calibrate the sensors at the indicated intervals.
Recalibration is not necessary if the flow sensor
has been unplugged only briefly.
Instructions for use Babylog 8000 plus SW 5.n
Monitoring
93