Operation
Information on pneumatic medication
nebulization
CAUTION
Reduced O
concentration
2
Air is used for medication nebulization. The
delivered O
concentration is therefore lower
2
than the set O
concentration.
2
If required, set a higher inspiratory O
concentration.
CAUTION
Insufficient medication nebulization
A medication nebulizer fault is not detected by
Babylog 8000.
– Check the correct functioning of the
medication nebulizer.
– Check whether aerosol is generated.
Medication nebulization may be used in all
ventilation modes.
The medication nebulizer nebulizes continuously.
The aerosol generated during expiration does not
reach the lungs, however.
72
O
concentration during nebulization
2
Air is used for medication nebulization. The
delivered O
the set O
If FiO
2
concentration using the nomogram.
2
Example:
With an inspiratory flow of 10 L/min and a desired
O
concentration of 80 Vol.%, the O
2
knob must be set to 90 Vol.% during medication
nebulization.
Instructions for use Babylog 8000 plus SW 5.n
concentration is therefore lower than
2
concentration.
2
must remain constant, set the O
2
-Vol.% setting
2

Advertisement