Dräger Babylog 8000 plus Instructions For Use Manual page 139

Intensive care ventilator for neonates
Hide thumbs Also See for Babylog 8000 plus:

Advertisement

Automatically set alarm limits
Airway pressure
Upper alarm limit for mechanical breaths
Upper alarm limit for expiration or CPAP
Lower alarm limit
Alarm limit for disconnection
O
concentration
2
Upper alarm limit
Lower alarm limit
Operating characteristics
Mains power supply
Mains power connection
Current consumption
At 230 V
At 110 V
Power consumption
Device fuses
Range 100 to 127 V
Range 230 to 240 V
Protection class
Ventilator
Proximal flow sensor (sensor connected)
Degree of protection
Gas supply
O
positive operating pressure
2
O
input flow
2
O
connection
2
Air positive operating pressure
Instructions for use Babylog 8000 plus SW 5.n
Description of alarm criteria, see page 143
Pinsp + 5 mbar (5 cmH
PEEP/CPAP + 4 mbar (4 cmH
PEEP/CPAP – 2 mbar (2 cmH
(Pinsp – PEEP) / 4 + PEEP
O
-Vol.% + 4 Vol.%
2
O
-Vol.% – 4 Vol.%
2
100 to 127 V 50/60 Hz
220 to 240 V 50/60 Hz
0.8 A
1.3 A
Approx. 140 W
T4AH250V IEC 60127-2/V 5x20 (2x)
T4AH250V IEC 60127-2/V 5x20 (2x)
Class 1
Type BF
or Type B
IP21
Protection from access with fingers and from solid
foreign objects with a diameter greater than
12.5 mm (0.47 in), protection from vertically drip-
ping water (only with storage tray fitted)
2.7 to 6 bar (270 to 600 kPa) (39 to 87 psi)
Up to 52 L/min
NIST
2.7 to 6 bar (270 to 600 kPa) (39 to 87 psi)
Technical data
O)
2
O)
2
O)
2
139