Assembly and preparation
Establishing potential equalization
Potential equalization allows electrical potential
differences between devices to be reduced.
Potential equalization does not replace the
protective earth connection.
During operation, the potential equalization
connectors must be readily accessible and the
connection must be able to be disconnected
without the use of tools.
Connecting the potential equalization
cable
1 Connect the potential equalization cable to the
potential equalization pin on the device.
2 Connect the potential equalization cable to a
hospital potential equalization connection (e.g.,
wall, ceiling supply unit, operating table).