Ventilation parameters
DCO
2
Continuous flow (set value)
Leakage rate
Spontaneous portion of minute vol-
ume
Factory-set default values:
Analog1: Flow
Analog2: Airway pressure
Setting the default values during operation
 Press the
/P
key.
The default values are used.
If a measured value exceeds the limits of the
scale, the voltage is limited to the maximum value
on the scale.
Instructions for use Babylog 8000 plus SW 5.n
Scale range
2
0 ... 200 mL
/s
2
0 ... 50 mL
/s
0 ... 125 L/min
0 ... 100 %
0 ... 100 %
–20 ... 20 L/min
–10 ... 90 mbar (–10 ... 90 cmH
Configuring the impulse output
The following signals are available:
– Triggered mandatory breath
– Mandatory breath
– Alarm
Default value: Mandatory breath
1 Open the dialog window by pressing the Cal.
Config. > Config > Com keys.
2 Press the Param key repeatedly until
highlighted.
3 Use the
Configuration
Measuring signal
range
→ 0 ... 10 V
→ 0 ... 10 V
→ 0 ... 10 V
→ 0 ... 10 V
→ 0 ... 10 V
O)
2
or
key to select the signal.