Maintenance
Preventive maintenance
WARNING
Risk of faulty components
Device failure is possible due to wear or
material fatigue of the components.
To maintain the function of all components,
this device must be inspected and serviced at
the intervals specified by the manufacturer.
Table of preventive maintenance intervals
Component
O
sensor
2
Cooling air filter
O-ring for expiratory
valve
Diaphragm of the expira-
tory valve
Lip seal for pneumatic
part
Battery for mains power
supply failure alarm
Flat seal on entry block
of gas connection
Filter on entry block of
gas connection
Time keeper RAM
Pressure reducers
124
Interval
When the O
sensor can
2
no longer be calibrated
Every 4 weeks
Every 12 months
Every 12 months
Every 12 months
Every 2 years
Every 2 years
Every 2 years
Every 2 years
Every 4 years
Every 6 years
WARNING
Risk of electric shock
Before performing any service work,
disconnect all electrical connections and gas
connections from power and gas supplies.
CAUTION
Risk of personal injury
If the glass on the LC screen is damaged, a
chemical liquid may escape.
– Avoid contact with the body.
– Immediately clean affected skin areas with
soap.
Measure
Replace, see page 42
Clean, see page 126
Replace
Replace
Replace
Replace
Replace
Replace
Replace
Replace
Basic overhaul of pres-
sure reducers
Instructions for use Babylog 8000 plus SW 5.n
Personnel responsible
Users
Users
Experts
Experts
Service personnel
Experts
Experts
Experts
Experts
Experts