Control and display unit
Layout of the control and display unit
The control and display unit consists of the display
panel (A) and the control panel (B). The display
panel includes the screen, the pressure display,
and keys with fixed and variable functions. The
control panel includes the setting knobs and keys
for ventilation.
Instructions for use Babylog 8000 plus SW 5.n
Setting knobs
A
B
A
D
The setting knobs are used to set the following
ventilation parameters:
A O
-Vol.%
2
B T
I
C T
E
D
Insp. Flow
E Pinsp
F PEEP/CPAP
The setting knobs, which can be set for the current
ventilation mode, are identified with green LEDs.
Setting ventilation parameters
 Turn the corresponding setting knob to the right
or to the left.
The set value is adopted immediately.
Turn the setting knobs for Pinsp and PEEP/CPAP
slowly. Otherwise, the set value will be displayed
incorrectly. After the next parameter has been set,
the display will be correct.
Operating concept
B
C
E
F
29