Technical data
Monitoring functions
Sound pressure level L
sured in accordance with IEC 60601-1-8:
Defined operating position: Front of the device at a
distance of 1 m (39 in) and a height of 1.5 m (59 in).
Sound pressure
Range for high-priority alarms according to
volume setting
Range for low- and medium-priority alarms
according to volume setting
Power supply failure alarm
Expiratory minute volume
Upper alarm limit alarm
Range
Lower alarm limit alarm
Range
Alarm suppression
Apnea alarm time
Alarm
Range
Respiratory rate (panting breathing)
Alarm
Range
Alarm delay time
Range
138
of alarm signals mea-
PA
63 dB (A) to 83 dB (A)
63 dB (A) to 82 dB (A)
53 dB (A)
MV
If the upper alarm limit has been exceeded
0.10 to 0.99 L/min in steps of 0.01 L/min
1.0 to 9.9 L/min in steps of 0.1 L/min
10 to 15 L/min in steps of 1 L/min
If the lower alarm limit has been exceeded
0 to 0.99 L/min in steps of 0.01 L/min
1.0 to 9.9 L/min in steps of 0.1 L/min
10 to 14 L/min in steps of 1 L/min
For the duration of the set alarm delay time
Apnoea time
If no breathing activity is detected
5 to 20 s in steps of 1 s
>20 s: monitoring is deactivated
Panting
If the alarm limit has been exceeded
20 to 200 bpm
<20 bpm: monitoring is deactivated
Alarm delay
Delays the following alarms:
– MV high
– VT low
Check settings!
0 to 30 s
Instructions for use Babylog 8000 plus SW 5.n