Dräger Babylog 8000 plus Instructions For Use Manual page 94

Intensive care ventilator for neonates
Hide thumbs Also See for Babylog 8000 plus:

Advertisement

Monitoring
O
monitoring
2
O
monitoring is always activated on Babylog
2
8000. The function cannot be deactivated.
Calibrating the O
2
CAUTION
Faulty calibration
If the quality of the oxygen from the central gas
supply system is insufficient, the calibration may
be faulty.
Calibrate the O
sensor with calibration gas
2
(100 % O
).
2
Performing calibration
1 Press the Cal. Config. key.
2 Press the O
-Cal key.
2
A
The device calibrates the O
message (A). O
-Cal (B) is displayed in the status
2
field.
The calibration is completed after approx.
5 minutes. O
-Cal (B) is no longer displayed in the
2
status field.
Hiding the message (A):
 Press the OK key.
94
sensor
B
sensor and displays a
2
Instructions for use Babylog 8000 plus SW 5.n