Dräger Babylog 8000 plus Instructions For Use Manual page 55

Intensive care ventilator for neonates
Hide thumbs Also See for Babylog 8000 plus:

Advertisement

Safety information
CAUTION
Malfunctions from condensation
Condensate may form when the device is moved
from a cold storage location to a warm
environment.
Only switch on the device when the condensate
has dried.
Switching on the ventilator
Conditions:
– The device has been reprocessed and
assembled ready for operation.
– The mains power supply and the gas supply
are connected.
CAUTION
Risk of unintentionally switching off the device
If the mains switch is not pressed until it locks
into position, but is only pushed under the
protective flap, the device may switch off. The
power supply failure alarm sounds when the
device switches off.
Press the mains switch until it locks in place.
Move the protective flap over the switch.
Instructions for use Babylog 8000 plus SW 5.n
1 Move the protective flap (A) on the rear of the
device to the side.
2 Press the mains switch (B) on the rear of the
device until it locks in place.
3 Move the protective flap over the mains switch.
Getting started
A
B
55