Getting started
2 Kink the expiratory breathing hose.
The visual alarm signal flashes red, the acoustic
alarm signal sounds, and one of the following
alarm messages is displayed:
– Hose kinked?
– Airway pressure high
Inspiration cancelled
3 Release the expiratory breathing hose.
4 Disconnect the connector on the Y-piece.
The visual alarm signal flashes red, the acoustic
alarm signal sounds, and one of the following
alarm messages is displayed:
– Airway pressure low
– Leak in hose system?
Check setting!
5 Set the PEEP/CPAP setting knob back to 0.
6 Reconnect the test lung.
Checking the breathing gas humidifier
Check the breathing gas humidifier in
accordance with the corresponding instructions
for use.
60
Device check immediately before using
the device on a patient
Carry out the device check immediately before
using the device on a patient in order to check the
operability of the device.
Conditions:
– Gas supply must be ensured.
– The expiration valve, breathing circuit, and flow
sensor must be correctly connected.
– The test lung must be connected.
Checking the breathing circuit for leaks
1 Switch the device on.
2 Press the OK key. The alarm message
Calibrate flow sensor! is no longer displayed.
3 Set the following ventilation parameters:
Parameter
Ventilation mode
Insp. Flow
Pinsp
4 Press the OK key.
5 Press and hold the man. Insp. key. The bar
display must read (80 ±2) mbar.
Calibrating the flow sensor
Calibrate the flow sensor, see page 95.
If calibration is successful, a confirmation is
displayed on the screen.
Checking the ventilation functions
1 Set the following alarm limits:
MV
0 L/min
MV
15 L/min
Instructions for use Babylog 8000 plus SW 5.n
Setting
CPAP
2
80