Intensive care ventilator for neonates
Hide thumbs Also See for Babylog 8000 plus:

Advertisement

High-frequency ventilation with CPAP
The high-frequency oscillations are continuously
being overlaid on the level of the mean pressure
(PEEP/CPAP).
Paw
PEEP/C
PAP
Flow
Instructions for use Babylog 8000 plus SW 5.n
High-frequency ventilation with IMV
Paw
Pinsp
PEEP/C
PAP
Flow
t
The high-frequency oscillations are overlaid
t
between the IMV breaths and the PEEP/CPAP
level in the expiratory time T
100 milliseconds before an IMV breath and
resume 250 milliseconds after the breath. The
pause following the breath is intended to allow
sufficient time to exhale and prevent air from being
trapped.
The
, Pinsp, PEEP/CPAP, and T
insp
have an effect on IMV breaths.
Mean airway pressure is somewhat higher than
with high-frequency ventilation with CPAP due to
the IMV breaths.
However, during high-frequency phases, the
pressure oscillates at the PEEP/CPAP level again.
Principles of operation
TI
. They end
E
settings
I
t
t
163