Download Print this page

Dräger Babylog 8000 plus Instructions For Use Manual page 57

Intensive care ventilator for neonates
Hide thumbs Also See for Babylog 8000 plus:

Advertisement

Charging the battery for the power supply failure alarm
WARNING
Absence of alarm during mains power supply
failure
If the battery is insufficiently charged, the
power supply failure alarm cannot sound in
the event of a mains power supply failure.
Charge the battery before the first use and
after being stored. Observe the charging time.
The battery is automatically charged when the
device is switched on.
Charge the battery for at least 30 minutes before
the first use and after being stored.
1 Make the following settings on the ventilator in
order to prevent alarms while the battery is
charging:
Parameter
O
-Vol.%
2
T
I
T
E
Insp. Flow
Pinsp
PEEP/CPAP
2 Switch the ventilator on.
3 Allow the charging time to expire.
Instructions for use Babylog 8000 plus SW 5.n
Setting
21
0.4 s
0.6 s
5 L/min
20 mbar (20 cmH
O)
2
3 mbar (3 cmH
O)
2
Getting started