Dräger Babylog 8000 plus Instructions For Use Manual page 121

Intensive care ventilator for neonates
Hide thumbs Also See for Babylog 8000 plus:

Advertisement

Overview
This chapter describes the necessary
maintenance measures required to maintain the
proper functioning of the medical device.
Maintenance measures must be performed by the
personnel responsible.
WARNING
Risk of infection
The responsible personnel may be infected
by pathogenic germs.
Disinfect and clean device or device parts
before any maintenance measures and also
before returning the medical device for repair.
Definition of maintenance concepts
Concept
Definition
Maintenance
All measures (inspection, preventive maintenance, repair) intended to maintain
and restore the functional condition of a medical device
Inspection
Measures intended to determine and assess the actual state of a medical device
Preventive mainte-
Recurrent specified measures intended to maintain the functional condition of a
nance
medical device
Repair
Measures intended to restore the functional condition of a medical device after a
device malfunction
Instructions for use Babylog 8000 plus SW 5.n
WARNING
Risk of electric shock
There are conducting components under the
housing cover.
– Do not remove the housing cover.
– Maintenance measures must be
performed by the personnel responsible.
Dräger recommends DrägerService for
repairs and complex maintenance tasks.
CAUTION
Malfunction
Soiled cooling air filters can compromise proper
operation of the device.
Replace the cooling air filters at regular intervals.
Maintenance
121