Dräger Babylog 8000 plus Instructions For Use Manual page 83

Intensive care ventilator for neonates
Hide thumbs Also See for Babylog 8000 plus:

Advertisement

Suppressing the acoustic alarm signal
The acoustic alarm signal can be suppressed for a
maximum of 2 minutes.
If another alarm occurs during this time, no alarm
signal sounds. However, the alarm message is
displayed on the screen.
If the fault triggering the alarm is not resolved after
2 minutes, the alarm signal sounds again.
 Press the
key.
The yellow LED on the key lights up.
Acknowledging alarm messages
After the fault triggering the alarm has been
resolved, the alarm tone is silenced. High-priority
alarm messages continue to be displayed and
need to be acknowledged.
 Press the OK key.
Instructions for use Babylog 8000 plus SW 5.n
Early reactivation of the alarm signal
 Press the
key.
The yellow LED on the key goes out.
Alarms
83