Suppressing the acoustic alarm signal
The acoustic alarm signal can be suppressed for a
maximum of 2 minutes.
If another alarm occurs during this time, no alarm
signal sounds. However, the alarm message is
displayed on the screen.
If the fault triggering the alarm is not resolved after
2 minutes, the alarm signal sounds again.
Press the
key.
The yellow LED on the key lights up.
Acknowledging alarm messages
After the fault triggering the alarm has been
resolved, the alarm tone is silenced. High-priority
alarm messages continue to be displayed and
need to be acknowledged.
Press the OK key.
Instructions for use Babylog 8000 plus SW 5.n
Early reactivation of the alarm signal
Press the
key.
The yellow LED on the key goes out.
Alarms
83