After reprocessing
Assembling the components
Installing the flow sensor insert
A
C
1 Slide the insert (A) into the flow sensor housing
(B) until it locks in place. Ensure that the two
markings are aligned.
2 Connect the plug (C) of the flow sensor cable
to the flow sensor.
Preparation for next use
1 Assemble and prepare the device for
operation, see chapter ''Assembly and
preparation'' on page 37.
2 Calibrate the flow sensor, see page 95.
3 Check readiness for operation, see chapter
''Getting started'' on page 54.
Instructions for use Babylog 8000 plus SW 5.n
B
Cleaning, disinfection and sterilization
119