Positive end-expiratory pressure or continuous
positive airway pressure
Range
Resolution
Accuracy
Trigger sensitivity
Range
Composition of gas
Performance characteristics
Control principle
Trigger delay time
Device compliance with breathing circuit without
breathing gas humidifier
Prerequisite:
– Inspiratory fraction: approx. 70 %
– Expiratory fraction: approx. 30 %
Device resistance with F&P 8411041 breathing cir-
cuit
Insp. Resistance
Exp. Resistance
Device resistance with F&P 8411153 HFV breathing
circuit
Insp. Resistance
Exp. Resistance
Device resistance with other breathing circuits:
Only use other breathing circuits if their resistance
values do not exceed the above-mentioned values
and the inspiratory fraction is approx. 70 %. Other-
wise the measurement of the airway pressure may
be impaired.
Instructions for use Babylog 8000 plus SW 5.n
PEEP/CPAP
0 to 25 mbar (0 to 25 cmH
0 to 10 mbar (0 to 10 cmH
(0.1 cmH
O)
2
>10 mbar (>10 cmH
O): 1 mbar (1 cmH
2
±1 mbar (±1 cmH
O) or 3 % of the measured value,
2
whichever is greater
Trig
1 to 10, corresponding to 0.02 to 3 mL
Breathing gas corresponding to the set O
tration, which can be set to NTPD or BTPS
Continuous flow, time controlled, pressure limited
O
is delivered by the integrated air/O
2
40 to 60 ms
<1.0 mL/mbar (<1.0 mL/cmH
12 mbar (12 cmH
O) at 30 L/min
2
5 mbar (5 cmH
O) at 30 L/min
2
6 mbar (6 cmH
O) at 30 L/min
2
4 mbar (4 cmH
O) at 30 L/min
2
Technical data
O)
2
O): 0.1 mbar
2
O)
2
concen-
2
mixer.
2
O)
2
133