Operation
When used without incubator:
1 Press the inlet port or the outlet port of the
medication nebulizer into one side of the hose
clip and the expiratory hose into the other.
2 Position the medication nebulizer vertically.
Preparing the medication nebulizer
 Fill the medication nebulizer in accordance with
the corresponding instructions for use.
Connecting the nebulizer hose
1 Insert the plug of the nebulizer hose into the
socket (E) until it locks in place.
Babylog 8000 starts nebulization.
2 Check whether aerosol is generated.
74
E
Increasing the O
concentration during
2
nebulization
1 The required O
concentration can be found on
2
the nomogram, see page 72.
2 Increase the O
concentration with the
2
O
-Vol.% setting knob.
2
Terminating medication nebulization
1 Retract the socket (F) so that the plug
disengages.
2 Remove any residual medication.
3 Remove the medication nebulizer and
reprocess it in accordance with the
corresponding instructions for use.
4 Set the O
-Vol.% setting knob to its original
2
value.
5 Install the flow sensor, see page 71.
6 Switch flow monitoring on, see page 71.
7 If a bacterial filter is used to protect the
expiratory valve, replace or remove the
bacterial filter.
8 If no bacterial filter is used, attach a
reprocessed expiratory valve, see page 42.
Instructions for use Babylog 8000 plus SW 5.n