Dräger Babylog 8000 plus Instructions For Use Manual page 136

Intensive care ventilator for neonates
Hide thumbs Also See for Babylog 8000 plus:

Advertisement

Technical data
Flow (t) waveform display on screen
Range
Resolution
Respiratory rate measurement
Respiratory rate
Range
Resolution
Accuracy
Inspiratory O
measurement
2
Sensor
concentration
Inspiratory O
2
Range
Resolution
Accuracy
T
0...90
136
–20 to 20 L/min
–2 to 2 L/min: 0.1 L/min
–5 to 5 L/min: 0.2 L/min
–10 to 10 L/min: 0.4 L/min
–20 to 20 L/min: 0.8 L/min
f
0 to 999 bpm
0 to 9.9 bpm: 0.1 bpm
10 to 999 bpm: 1 bpm
±1 bpm
Fuel cell integrated in device
Measurement is taken in the inspiratory flow.
The sensor is not barometrically pressure compen-
sated. In the event of significant air pressure fluctu-
ations within a calibration interval (24 hours), it may
be necessary to perform a manual calibration.
FiO
2
18 to 100 Vol.%
1 Vol.%
±3 Vol.% based on the compressed gases O
Air
<65 s at 1 L/min
<10 s at 30 L/min
Instructions for use Babylog 8000 plus SW 5.n
and
2