Operation
Analog and digital interface
The analog and digital interface is available as an
option and can be used for the following functions:
– Analog output of measured values
– Output of protocols
– Communication with the patient monitor or
personal computer, e.g., with the BabyView PC
program for graphic and numerical
representation of ventilation parameters.
CAUTION
Incorrectly transferred data
All transferred data is for informational purposes
only and must not be used as basis for diagnostic
or therapeutic decisions.
Check the displays on Babylog 8000. The
MEDIBUS interface is not intended for use with a
"distributed alarm system" according to
IEC 60601-1-8:2012.
Set up any connected devices in the same room
as Babylog 8000 in accordance with
IEC 60601-1-1. A minimum distance of 1.5 m
(59.1 in) to the patient must be maintained.
Analog output ports:
Each analog output port outputs one of the
available measured values.
RS232 interface:
Data is output via the RS232 interface to a patient
monitor or personal computer, or the following
information is output to a printer:
– Protocol
– Content of the trend memory
– Measurement waveforms
Or
– Data is transmitted to a patient monitor or
personal computer
76
Impulse output port:
The impulse output port
following events:
– Triggered mandatory breath
– Mandatory breath
– Alarm
Data output
Technical requirements can be found in chapter
''Technical data'' on page 130.
For information on configuring the interfaces, see
chapter ''Configuring the data interfaces'' on
page 101.
Analog output of measured values
One measured value, e.g., VT, Paw, Flow, can be
output through each of the outputs Analog1 (A)
and Analog2 (B).
Voltage range: 0 to 10 V
Analog 1
Analog 2
A
1 Connect the printer (internal resistance ≥
1 MΩ) using the cable 8306487.
2 Select the measuring signal and scale range,
see page 101.
Instructions for use Babylog 8000 plus SW 5.n