2 Disconnect the probe of the O
gas hose from the wall terminal unit of the
central gas supply system.
The visual alarm signal flashes red, the acoustic
alarm signal sounds, and the alarm message O
pressure low is displayed.
3 Plug the probe into the wall terminal unit of the
central gas supply system.
The visual alarm signal goes out, the acoustic
alarm signal is silenced, and the alarm message is
no longer displayed.
Checking the compressed air alarm:
1 Disconnect the probe of the compressed air
gas hose from the wall terminal unit of the
central gas supply system.
The visual alarm signal flashes red, the acoustic
alarm signal sounds, and the alarm message
Medical air low is displayed.
2 Plug the probe into the wall terminal unit of the
central gas supply system.
The visual alarm signal goes out, the acoustic
alarm signal is silenced, and the alarm message is
no longer displayed.
Checking the IPPV ventilation mode
1 Set the following ventilation parameters:
Parameter
Ventilation mode
O
-Vol.%
2
Insp. Flow
T
I
T
E
Pinsp
PEEP/CPAP
2 On the main screen, press the Meas key.
3 Press the Paw key.
Instructions for use Babylog 8000 plus SW 5.n
compressed
2
2
Setting
IPPV/IMV
21
10
0.4
0.6
20
0
The displayed values must be within the following
ranges:
Measured
Permissible range
value
Peak
18 to 22 mbar (18 to 22 cmH
Mean
6 to 10 mbar (6 to 10 cmH
PEEP
–1.5 to 1.5 mbar
(–1.5 to 1.5 cmH
Checking the PEEP
1 Set the PEEP/CPAP setting knob to 10.
2 Press the OK key.
3 On the main screen, press the Meas key.
4 Press the Paw key.
The displayed values must be within the following
ranges:
Measured
Permissible range
value
Peak
18 to 22 mbar (18 to 22 cmH
Mean
12 to 16 mbar (12 to 16 cmH
PEEP
8 to 12 mbar (8 to 12 cmH
Checking the alarm limits
Checking apnea monitoring:
 Set the CPAP ventilation mode.
After a maximum of 30 seconds, the visual alarm
signal flashes red, the acoustic alarm signal
sounds, and one of the following alarm messages
is displayed:
– Apnoea
– MV low
Checking the alarm limits for airway pressure:
1 Set the IPPV/IMV ventilation mode.
Getting started
O)
2
O)
2
O)
2
O)
2
O)
2
O)
2
59