Cleaning, disinfection and sterilization
Dismantling the Y-piece and flow sensor
1 Remove the plug of the flow sensor cable from
the flow sensor.
2 Loosen and remove the plug of the flow sensor
cable from the rear of the device.
3 Remove the Y-piece from the breathing hoses.
4 Remove the ISO 15 flow sensor from the Y-
piece.
5 Gently press the knobs on both sides of the
flow sensor insert while pulling the insert out of
the flow sensor.
Reprocessing the Y-piece and flow sensor
 Reprocess the insert of the flow sensor
immediately after every use, see page 117.
 Reprocess the ISO 15 flow sensor housing and
the Y-piece in accordance with the
reprocessing list, see page 117.
 Reprocess the Y-piece flow sensor in
accordance with the respective reprocessing
instructions and the reprocessing list, see
page 117.
112
Dismantling the pneumatic medication
nebulizer
D
1 Remove the nebulizer hose (C) from the
medication nebulizer (B) and from the
nebulizer port on the device.
2 Remove the medication nebulizer (B) from the
breathing circuit.
3 Remove the corrugated hose of the breathing
circuit (D) from the inlet port.
4 Remove the corrugated hose (A) from the
outlet port.
5 Disassemble the medication nebulizer in
accordance with the corresponding instructions
for use.
Reprocessing the medication nebulizer and the
parts for adaptation
 Reprocess the individual parts of the
medication nebulizer in accordance with the
corresponding instructions for use.
 Reprocess the parts for adaptation in
accordance with the reprocessing list, see
page 117.