Flow monitoring
Calibrating the flow sensor
WARNING
Risk of fire
Residual vapors of easily flammable
disinfectants (e.g., alcohols) and deposits
that were not removed during reprocessing
can ignite when the flow sensor is in use.
– Ensure particle-free cleaning and
disinfection.
– After disinfection, allow the flow sensor to
air for at least 30 minutes.
– Before inserting the flow sensor, check for
visible damage and soiling such as
residual mucus, medication aerosols, and
particles.
– Replace flow sensors when damaged,
soiled, or not particle-free.
Information on calibration
The sensor type being used must be selected in
order to optimize the precision of the
measurement:
ISO
ISO 15 flow sensor (8411130)
Y
Y-piece flow sensor (8410185)
The applicable reference conditions must be
selected:
NTPD
Ambient temperature 20 °C (68 °F), air
pressure 1013 mbar (or hPa or
cmH
O), dry gas
2
BTPS
Body temperature 37 °C (98.6 °F), air
pressure 1013 mbar (or hPa or cmH
plus mean airway pressure 10 mbar (or
hPa or cmH
2
gas
Instructions for use Babylog 8000 plus SW 5.n
O), water-vapor-saturated
Preparing calibration
1 Press the Cal. Config. key.
2 Press the Sensor key (A).
3 Select the Flowsensor line by pressing the
key.
4 Select the sensor type being used by pressing
the
key or the
5 Select the Ref. cond. line by pressing the
key.
6 Select the reference conditions using the
key or the
7 Press the
B
O)
2
8 Press the
-Cal
Monitoring
A
key.
key.
key.
key (B).
95