4.5 Electrical installation

4.5.1 Check and configure the pump

CAUTION:
Isolate the power supply before changing the voltage. Failure to configure the pump
electrical supply correctly can result in damage to the pump.
Ensure that the voltage shown on the voltage indicator (refer to
the motor cover corresponds with the electrical supply voltage. If not, change the
configuration of the pump motor to match the supply voltage using the following
procedure.
1.
Undo the two screws and lift off the voltage indicator moulding.
2.
Rotate the voltage indicator moulding so that the correct voltage is uppermost.
3.
Replace the two screws.

4.5.2 Connect the pump to the electrical supply

WARNING:
Ensure that the XDS pump electrical installation conforms to local and national safety
requirements. It must be connected to a suitably fused and protected electrical
supply with a suitable earth (ground) point.
WARNING:
If using an earth leakage device, for example, an RCD, use a 30 mA rated unit at
minimum to avoid tripping during start up.
WARNING:
The pump will automatically restart after restoration of the power supply following
power failure.
For recommended regional supply protection and cordsets, refer to
page 15.

4.6 Inlet and outlet connections

WARNING:
If pumping dangerous gases or vapours, connect the exhaust to a suitable treatment
plant to prevent the discharge of dangerous gases and vapours to the surrounding
atmosphere.