31 Help and error messages
31.3.3 Control panel related error messages
31.3.4 Instrument related error messages
268
Planmeca Compact i
Power supply related error messages
ERROR
EXPLANATION
CODE
E16.3
Scaler electronics slot does not match with scaler hose
connection.
Check scaler electronics connection.
E16.4
Other instrument than scaler is connected to slot where scaler
electronics is.
Check scaler electronics connection.
E17
Power Supply related error messages.
E18
Power Supply related error messages.
Control panel related error messages
ERROR
EXPLANATION
CODE
E19.1
Control panel in branch 1 is of new type.
Update MCB software.
E19.2
Control panel in branch 2 is of new type.
Update MCB software.
E20.1
Control panel 1 button stuck.
Control panel button is stuck during self test.
E20.2
Control panel 2 button stuck.
Control panel button is stuck during self test.
E21
Control panel related error messages.
E22
Control panel related error messages.
Instrument related error messages
ERROR
EXPLANATION
CODE
E23
Polymerisation light bulb has blown or is not properly attached
to socket.
E23.1
Instrument overtemperature error.
Check the instrument handpiece and let the instrument cool
down.
E23.2
Micromotor error; phase missing.
Check that the micromotor and hose are connected properly.
E23.3
Instrument undervoltage error.
E23.4
Instrument power protection error.
E23.5
Instrument EEPROM error.