31 Help and error messages
Help messages
HELP
MEANING
CODE
H 23
Automatic/manual chip blow cannot be used
when sterile water function is selected.
H 24
Programming of selected setting is disabled. The programming of the selected setting is
H 25
Planmeca Compact e help message:
Patient chair cannot be driven into
automatic chair position.
H 26
Drive chair upwards, then lower backrest. If
patient chair is attached to unit, chair's
automatic positions are not switched on.
H 27
Drive backrest upwards, then lower chair. If
patient chair is attached to unit, chair's
automatic positions are not switched on.
H 28
Planmeca Compact WE's help message.
H 29
Drive chair downwards, then lower backrest.
If patient chair is attached to unit, chair's
automatic positions are not switched on.
H 30
Close water tap. Fill container with
disinfectant up to groove and attach to unit.
H 31
Switch unit off, leave disinfectant in unit
overnight (min. 8 h).
H31.1
Remove container, fill it up to groove with
water and attach to unit.
H31.2
Attach container filled with water.
H 32
Container flushing completed.
H 33
Open water tap.
H 34
No instruments selected when starting
instrument flushing or waterline cleaning.
258
Planmeca Compact i
ACTION OR EXPLANATION
disabled.
Contact your Planmeca dealer.
Contact your Planmeca dealer.
Drive the chair upwards before lowering the
backrest. If this help message is shown
while the patient chair is attached to the
unit, it means that you have forgotten to
switch on the chair's automatic positions. To
clear the message, put the operation mode
switch on the cuspidor side of the chair base
in the "AUTO" position (switch up).
Drive the backrest upwards before lowering
the chair. If this help message is shown
while the patient chair is attached to the
unit, it means that you have forgotten to
switch on the chair's automatic positions. To
clear the message, put the operation mode
switch on the cuspidor side of the chair base
in the "AUTO" position (switch up).
Connect the lifter cable.
Drive the chair downwards before lowering
the backrest. If this help message is shown
while the patient chair is attached to the
unit, it means that you have forgotten to
switch on the chair's automatic positions. To
clear the message, put the operation mode
switch on the cuspidor side of the chair base
in the "AUTO" position (switch up).
When unit is turned back on, disinfection
procedure continues.
Remove container from unit.
Select at least one instrument and syringe
(remove from holder, bend the instrument
arms).
User's manual