27 Flushing and cleaning programs
Results
27.6 Waterline cleaning
27.6.1 Introduction
214
Planmeca Compact i
After interrupting the cleaning program you can start it again by filling the
dosing cubic and pressing the green start button. The program will start
from the beginning.
1. Dosing cubic
2. Suction tube cleaning holder
3. Start button (green)
4. Cancel button (yellow)
9. When the cleaning program is completed, help message H 42 is
displayed. Return the suction tubes to the suction holder and close the
suction cleaning cover.
10. If suction disinfectant is left in the dosing cubic after the cleaning cycle,
clean the cubic under running water.
The dental unit is now ready for normal operation.
NOTE
Before attaching the handpieces back on the suction tubes, wipe the
identification bushings (1) with Planmeca approved surface disinfectant.
Waterlines are a breeding ground for biofilm. Biofilm may contain bacteria
that are harmful to the patient and to the dental team. This is why the
waterlines of the dental unit should be cleaned once a week with a
disinfectant. The solution is left in the unit overnight and the waterlines of the