31 Help and error messages
Help messages
HELP
MEANING
CODE
H 47
Suction disinfectant pressure is low. (In
automatised suction cleaning: Fill suction
disinfectant container).
H 47.1
Suction disinfectant level low. Fill suction
disinfectant container.
H 47.2
Suction disinfectant container filled.
H 48
WMS must be set to Bottled water off and
Bottle on -modes before starting suction
cleaning.
H 50
No cup in cup holder. Sensor does not
recognise cup when pressing cup fill button.
H 51
Bowl is above patient chair when driving
chair up.
H 52
Patient chair can not be driven upward
because safety switch of side delivery arm
is released.
H 53
Patient chair can not be driven downward
because safety switch of side delivery arm
is released.
H 54
Cup is inserted in cup holder.
H 55
Disinfectant container attached. To start
waterline cleaning, select instruments and
press Waterline cleaning button.
H 57
Vision action failed. Select patient in
Romexis.
H 60
WMS: Container almost empty in Domestic
water supply -mode.
H 61
WMS: Container almost empty in Bottled
water -mode.
H 62
WMS: Container switch is turned off.
H 63
Cuspidor door is open.
H 64
Fill container with disinfectant over upper
sensor.
H 65
Fill container with water over upper sensor.
H 66
Wait 8 hours or switch unit off for the night.
H 67
Instrument flush cancelled. Return
instruments to instrument console.
H 68
Romexis connection lost.
H 69
Unable to create a new user. Maximum
amount of local users exceeded.
260
Planmeca Compact i
ACTION OR EXPLANATION
Bottled water off = domestic water is used
Bottle on = bottle is used; domestic water is
run through the bottle
Place a cup on the cup holder. Or change
the setting of service mode n 115.
If the unit has a bowl, move the bowl to the
rest position. If the unit has an OP delivery
arm safety switch, check that the OP
delivery arm is not blocking the chair
movement.
Remove the cup and make sure the cup fill
tube is above the bowl.
Select patient in Planmeca Romexis before
performing Solanna Vision action.
Reduce water use for a moment.
Reduce water use and fill the container as
soon as possible.
Switch on the container switch.
Close the door.
Make sure the Ethernet cable is attached
and Romexis is running.