25 Programming

25.4 Instrument settings

25.4.1 Instrument speed/power limit

About this task
Steps

25.4.2 Instrument spray

Steps
176
Planmeca Compact i
When you have signed in to the dental unit with your own user name, all
changes you make to the instrument settings and then save, are saved to
your personal settings. This means that whenever you sign in to the dental
unit, you will use your last saved instrument settings.
NOTE
The instrument speed/power limit does not affect the air driven instruments
for which quickstart has been selected.
1. Activate the instrument.
2. Press Program.
3. Press Instrument speed/power limit.
When you are programming the turbine, press Turbine speed/power
limit.
A pop-up window opens.
4. For W&H and Minetto brushless micromotors only: Press %/rpm to
select whether you want to adjust the instrument speed/power limit in
percentage points or in rpm. The appearance of the button in the
instrument view changes accordingly.
5. Adjust the instrument speed/power limit.
The minimum speed/power limit is 5 or 10% (depending on the
instrument) and the maximum is 95%. The adjustment step is 5.
If the preset value is, for example, 50%, the foot control pedal movement
controls the speed/power between 5 - 50% instead of the normal range
of 5 - 100%.
6. Press OK.