Planmeca Compact i Touch v2 User Manual page 269

Table of Contents

Advertisement

Help messages
HELP
MEANING
CODE
H 35
Instruments are locked for safety reasons.
H35.1
Waterline cleaning cancelled or failed.
Remove container.
H 36
Flushing ready. Return instruments to
instrument console.
H36.1
Instrument flushing completed, but water
flow was not detected for one or more
instruments.
H36.8
Low water pressure during final flushing.
H 37
Waterline cleaning has been interrupted.
H 38
Low water pressure during final rinsing.
H 39
Syringe misplaced or its flow not detected in
waterline cleaning.
H 40
Too many Bien-Air MX brushless
micromotors.
H 41
Insert suction tubes to suction tube cleaning
holder, add disinfectant and press green
start button (upper button).
H 41.1
Insert suction tubes to suction tube cleaning
holder.
H 41.2
Remove suction tubes from holder to start
suction cleaning.
H 42
Suction cleaning completed. Return suction
tubes to suction holder and close suction
cleaning cover.
H 42.1
Suction cleaning completed. Return suction
tubes to suction holder.
H 43
Suction cleaning cancelled. Return suction
tubes to suction holder.
H 44
Incorrect suction tube amount detected.
Return them to suction holder, place them
back to suction tube cleaning holder and
start cleaning.
H 45
Set WMS to Domestic Water Supply -mode
and wait until WMS bottle is full. Then start
suction cleaning.
H 46
Suction use is forbidden or there are no
suction tubes in dental unit configuration.
User's manual
31 Help and error messages
ACTION OR EXPLANATION
Select instruments, place them in flushing
holder, and press Instrument flushing button
to start instrument flushing.
Make sure water tap is open. Select
instruments, place them in flushing holder
and press Instrument flushing button to start
instrument flushing.
Also bend and release syringe's instrument
arm once.
Run instruments for a while with spray water
to ensure instrument flushing is sufficient.
Perform long instrument flushing to ensure
instruments are flushed.
When waterline cleaning is finished, perform
long instrument flushing to ensure
instruments are flushed.
Return syringe to console. Then, select
syringe (bend instrument arm) again and
place it carefully in flushing holder.
A maximum of two is allowed.
Suction cleaning can not be used when
WMS is set to Bottled Water -mode.
The value of service mode n.83 is 0.
Planmeca Compact i
259