7 For your safety
CAUTION
Drops of water on the touch display might disturb the functioning of the
control panel.
CAUTION
Lock the control panel from the Maintenance window before using a table-
top instrument.
CAUTION
The dental unit must not be used simultaneously with the intraoral X-ray unit
Planmeca ProX.
CAUTION
The light source of the operating light may cause retinal injury if viewed upon
directly.
Protect the patient's and dental treatment staff's eyes with protective glasses
that block high-energy visible light (HEV light), or limit the direct exposure to
10 minutes.
CAUTION
Switch off the unit before using an electrosurgical knife.
CAUTION
The use of the electrosurgical knife may affect the function of an implanted
pacemaker or defibrillator. Please refer to the manufacturer's own
documentation.
CAUTION
Do not use the scaler or the polymerisation light on patients with cardiac
pacemakers. The instrument can cause disturbance on the pacemaker's
function.
CAUTION
Electromagnetic interference between the equipment and other devices can
occur in very extreme conditions. Do not use the equipment in close
conjunction with sensitive devices, or devices creating high electromagnetic
disturbances.
CAUTION
Do not use the equipment in close conjunction with anaesthetic gas or in
highly oxygenated environments (oxygen content >25%).
CAUTION
Before using the dental unit, ensure that the instruments have been properly
flushed and that the suction tubes as well as the dental unit's waterlines
have been cleaned as instructed in this manual.
CAUTION
If the drain is blocked, the dental unit might overflow with contaminated
water and excessive water could flow onto the floor. Shut down the unit and
contact your property's caretaker and your Planmeca dealer.