28 Cleaning and disinfection

28.5.4 Weekly cleaning procedures

28.5.4.1 During working day
About this task
Steps
236
Planmeca Compact i
2. Place the rinsing bottle (for example OroCup) on a flat surface (table or
floor).
3. Remove the suction handpieces from their holders and push them onto
the inserts inside the cap of the rinsing bottle. Allow the bottle to empty.
4. Return the suction tubes back to the holder right after OroCup has been
emptied. Do not use suction only for air.
5. Replace the exchangeable parts (filters etc.) the next morning.
Clean the suction system with Dürr MD 555 cleaner 1 - 2 times a week to
prevent the buildup of deposits in the suction system, especially if you are
using dental air polishers.
This procedure is mandatory for dental units with a Dürr amalgam separator
or Dürr VS/A separator.
For more information on Dürr MD 555 cleaner, please visit
http://www.duerrdental.com.
1. Pour 50 ml of Dürr MD 555 cleaner into a rinsing bottle (for example
OroCup). Add 1 l water and mix well.
2. Place the rinsing bottle on a flat surface (table or floor).
3. Remove the suction handpieces from their holders and push them onto
the inserts inside the cap of the rinsing bottle.
4. Remove the suction handpieces from the rinsing bottle when there is 250
ml of the solution left in the bottle.
User's manual