Suction System; In The Morning; After Each Patient - Planmeca Compact i Touch v2 User Manual

Table of Contents

Advertisement

28 Cleaning and disinfection

28.5 Suction system

28.5.1 In the morning

Steps

28.5.2 After each patient

Steps
Results
234
Planmeca Compact i
NOTE
Dry the suction tube holders and the tube bushings properly after cleaning.
Wet surfaces might disturb suction tube recognition.
1. Rinse each suction tube with 0.5 litres of water by slowly aspirating water
and air through the suction handpieces.
2. Wipe the suction handpieces with Planmeca approved surface
disinfectant.
1. Remove the used aspirating tips.
2. For hygienic and operational reasons, empty one glass of water (100 -
200 ml) with each suction tube by aspirating water and air through the
suction handpiece. Do this even if only the saliva suction tube has been
used.
3. Wipe the suction handpieces with Planmeca approved surface
disinfectant.
4. Wipe the suction holder / Flexy-holder, including the tablet holder, with
Planmeca approved surface disinfectant.
NOTE
Do not use a spray disinfection solution for the suction arms and holders.
NOTE
Clean the tablet according to the instructions given by the manufacturer.
User's manual

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Compact i Touch v2 and is the answer not in the manual?

Questions and answers

Table of Contents

Save PDF