Help messages
HELP
MEANING
CODE
H 92
Help (Stop) button pressed.
H 93
This button has no function at the moment
because there is no motorised light.
H 94
Clock calibration failed. Maximum
calibration range exceeded.
H 95
Patient chair automatic position cannot be
reached because legrest is unlocked or
armrests are blocking movement.
H 96
PlanID tag already in use.
H 96.1
Different user already logged in.
H 97
Failed to assign PlanID tag to user.
H 98
User not found.
H 98.1
User login failed.
H 98.2
RFID unknown.
H 99
Unit has been switched off without
performing long flushing after water has
been used. Perform long flushing.
HE200 Detected new software update. Confirm to
install?
HE201 Detected new software update SD-card.
Confirm to install 5.3.0.9.R ?
HE210 Software update. Installing to Control panel
bootloader sw. DO NOT TURN POWER
OFF.
HE212 Software update. Installing to Control panel
application sw. DO NOT TURN POWER
OFF.
HE214 Software update. Installing to Control panel
application sw. DO NOT TURN POWER
OFF.
HE216 Software update. Downloading from server.
DO NOT TURN POWER OFF.
HE218 Software update. Installing to Main PCB.
DO NOT TURN POWER OFF.
END
Flushing completed. Remove syringe from
flushing holder and return it to console. If
syringe is removed, return it to flushing
holder.
DOOR
Cuspidor door is open.
User's manual
31 Help and error messages
ACTION OR EXPLANATION
PlanID tag assigned to another user.
The unit has rejected a login request
because a different user is already logged
in.
Check Romexis connection.
User not found with this PlanID tag. Create
a new user or select an existing Romexis
user and assign the PlanID tag.
Retry login or reboot the unit. If problem
persists, contact service.
The RFID is unknown to the system.
This help message can be disabled, contact
your Planmeca dealer.
Do not turn the power off while the software
is being updated.
Do not turn the power off while the software
is being updated.
Do not turn the power off while the software
is being updated.
Do not turn the power off while downloading
software from the server.
Do not turn the power off while the software
is being updated.
Close the door.