25 Programming
25.4.4 Automatic chip blow
About this task
Steps
25.4.5 Instrument light
Steps
178
Planmeca Compact i
NOTE
If you disable the sterile water mode for an instrument that uses sterile
water, the flow rate for sterile water 1 and 2 automatically changes to
0%.
5. Press OK.
The pop-up window closes.
NOTE
The type of manual chip blow does not depend on the type of the automatic
chip blow, but is dry as a factory default. The manual chip blow can be set to
be dependent on programming, contact your Planmeca dealer.
1. Activate the instrument.
2. Press Program.
3. Press Chip blow.
A pop-up window opens.
4. Adjust the flow rates for water and air.
The minimum value of both parameters is 0 (no flow) and maximum 100.
The adjustment step is 1 in the value range 0 - 35% and 5 in the value
range 35 - 100%.
5. Press OK.
The pop-up window closes and the new values are displayed on the
control panel.