7 For your safety
NOTE
If using air abrasives, please consult instructions for use given by the air
polisher manufacturer. Make sure that adequate flushing of dental unit is
performed immediately after using air polishers.
NOTE
The air used by the dental unit instruments must be dry, clean and oil-free.
NOTE
Never place heavy objects or containers of liquid on any part of the unit or
hang objects from the unit's arm structures.
NOTE
Care should be taken when other movable equipment is used in conjunction
with the dental unit.
NOTE
The dental unit shall only be connected to a trusted private network (and not,
for example, the Internet).
NOTE
EMC requirements have to be considered, and the equipment must be
installed and put into service according to the specific EMC information
provided in the accompanying documents.
NOTE
Portable and mobile RF communications equipment can affect the dental
unit.
NOTE
External equipment intended for connection to signal input, signal output or
other connectors, shall comply with relevant IEC standard (e.g. IEC 60950
for IT equipment and the IEC 60601 series for medical electrical equipment).
In addition, all such combinations - systems - shall comply with the IEC
60601-1 standard. Equipment not complying to IEC 60601-1 shall be kept
outside the patient area.
Any person who connects external equipment to signal input, signal output
or other connectors has formed a system and is therefore responsible for the
system to comply with the requirements of IEC 60601-1. If in doubt, contact
a qualified technician or your local representative.