CAUTION
Endodontic treatment must only be performed using the Morita TORX
micromotor's a1 - a3 presets and the Morita CA-10RC-ENDO 10:1
handpiece.
CAUTION
Always check the preset settings before using the micromotor.
The table below lists the factory presets for the Morita TORX micromotor.
These presets can be modified, see section "Modifying preset values" on
page 184.
The tolerance for RPM and torque values is +/- 5%.
NOTE
The torque limit value depends on the handpiece type.
Preset 1 - 3
Preset a1
Spray 1:
Disabled
Water: 50
Air: 80
Spray 2 (Dry):
Water: 0
Air: 80
Default: Spray
off
Available
N/A
Enabled
Disabled
On/Off
Disabled
Water: 0
Air: 60
Default: Chip
blow on
Can be
N/A
configured to
be used from
foot control
Default: On
Forward/
Forward/
Reverse
Reverse
Default:
Default:
Forward
Forward
rotation
rotation
Preset a2
Preset a3
Disabled
Disabled
N/A
N/A
Disabled
Disabled
Disabled
Disabled
N/A
N/A
Forward/
Forward/
Reverse
Reverse
Default:
Default:
Forward
Forward
rotation
rotation
Preset t1 - t3
Spray 1:
Water: 50
Air: 80
Spray 2 (Dry):
Water: 0
Air: 80
Default: Spray
off
Available
Enabled
On/Off
Water: 0
Air: 60
Default: Chip
blow on
Can be
configured to
be used from
foot control
Default: On
Forward/
Reverse
Default:
Forward
rotation
User's manual