Help messages
HELP
MEANING
CODE
H 70
Brushless micromotor is stuck or broken.
H 71
Romexis user ID already in use.
H 72
Incompatible Romexis version.
H 72.1
Incompatible Romexis version
H 73
Failed to copy user profile from Romexis.
H 73.1
Failed to save user settings.
H 74
PlanID reader missing.
H 75
Yearly maintenance in <#> days.
H 76
Yearly maintenance performed.
H 79.1
Serial number chip missing.
H 80
Headrest length safety feature activated
when driving headrest. Check that nothing is
blocking headrest movements.
H80.1
Headrest length drive limit activated when
driving headrest.
H 81
Headrest angle B safety feature activated
when driving headrest. Check that nothing is
blocking headrest movements.
H81.1
Headrest angle A drive limit activated when
driving headrest.
H 82
Connecting to wireless foot control was
successful.
H 83
Foot control handle pressed down.
H 84
Connecting to wireless foot control failed.
User's manual
31 Help and error messages
ACTION OR EXPLANATION
Check the ID, select another one, or allow
rebind in Romexis.
PlanID is not supported with this version of
Romexis.
User settings for currently active instrument
can not be used. Romexis version does not
support user instrument settings or presets
for currently active instrument. Check the
instrument settings before using instrument
and update Romexis.
PlanID reader is either missing or not
functional.
Contact service for yearly maintenance.
The serial number chip is missing. Contact
service.
Check that nothing is blocking the headrest
movement. The headrest can be driven
normally after the possible obstruction has
been removed.
Check that nothing is blocking headrest
movements.
Faulty length motor, motor cable,
nonlubricated slider rail or position
potentiometer may cause this problem.
Check that nothing is blocking the headrest
movement. The headrest can be driven
normally after the possible obstruction has
been removed.
Faulty angle B motor, motor cable or
position sensor may cause this problem.
Check that nothing is blocking the headrest
movement.
Faulty angle A motor, motor cable or
position sensor may cause this problem.
Connection was established with the
requested foot control.
Release the foot control handle.
Connection with the requested foot control
could not be established. Make sure the foot
control is active (connect to a charger, if
possible), is in the vicinity of the unit, and try
again.
Planmeca Compact i
261