28 Cleaning and disinfection
Saliva suction handpiece
28.6 Planmeca WEK water disinfection system
28.7 Planmeca ProX
238
Planmeca Compact i
If necessary, the handpiece can be completely disassembled for more
thorough cleaning. The handpiece parts can be disinfected in a washer-
disinfector at 93°C, then optionally autoclaved at 134°C.
The saliva suction handpiece can be disassembled for cleaning by pulling
out the end of the handpiece and removing the core (2) and the adjuster (1).
The handpiece parts can be disinfected in a washer-disinfector at 93°C, then
optionally autoclaved at 134°C.
For information on how to use the Planmeca WEK water disinfection system,
Planmeca WEK water disinfection system user's manual .
see
For information on how to clean the Planmeca ProX X-ray unit, see
Planmeca ProX user's manual .