Calibration; Run The Test/Measurement - Keysight PD1000A Startup Manual

Power device measurement system for advanced modeling
Hide thumbs Also See for PD1000A:
Table of Contents

Advertisement

On State and Off State S-Parameter Characterization
2 (Optional) To save set-up time, you can save all configuration settings and

Calibration

Calibrate the ENA with the appropriate PD1000A Test Fixture connected. Use the
Calibration Standards supplied with the test fixture. Refer to the PD1000A Test
Fixture User Guide or the PD1000A Help file for calibration details.

Run the Test/Measurement

1 Click the Start button (
74
-- Drain Voltage (V): Enter the voltage for SMU Channel 2 (rear panel) of the
B2902A SMU. Typically, this connects to the drain of the FET or
collector of the IGBT.
-- Drain Current Limit (A): Enter the Current Limit for SMU Channel 2 (rear
panel) of the B2902A SMU.
general settings to a file that you can re-use to skip the configuration process
the next time you test a DUT that uses the same settings.
a Click on Settings on the main screen.
b Select Save Settings to File. This displays a file save dialog.
c Use the dialog to browse to the directory where you want to store the
settings.
d Type a file name for the settings file using one of the following formats:
-- IVCV.kts - for IV-CV tests
-- ONSTATE.kts - for on-state S-parameter tests
-- OFFSTATE.kts - for off-state S-parameter tests
e Click Save. This saves your current settings into a Keysight Test Settings file
(.kts).
– You can stop an active measurement by clicking the Stop (
Running the S-Parameter Characterization
) to start a measurement.
Keysight PD1000A Power Device Measurement System Startup Guide
) button.