On State and Off State S-Parameter Characterization
3 From the E5080A ENA Analyzer, connect the "N" Adapters and SMA cables to
SMA Cable to
ENA Port 1
Figure 34 E5080A ENA with Bias Ts and Test Fixture (used for on-state and off-state
S-parameter measurements)
4 Ensure that the PC can communicate with the B2902A SMU and the E5080A
68
the two PD1000A Bias T networks as shown in Figure 34 below:
Device Test Fixture (SMD shown)
ENA using the Keysight Connection Expert utility in the IO Libraries Suite.
To communicate with the B2902A SMU and the E5080A ENA,
Keysight IO libraries must be installed on your external controller.
To obtain the latest version of Keysight IO libraries, go to
www.keysight.com/find/IOSuite.
Every E5080A ENA is shipped with the Keysight IO libraries
installed. Do NOT upgrade the Keysight IO libraries on the ENA.
Unexpected results may occur.
Keysight PD1000A Power Device Measurement System Startup Guide
Test Requirements
SMA Cable to
ENA Port 2