IV and CV Device Characterization
Setup the CV to IV Switch.
1 Specify the HVSMU slot. (See Figure 15 above and refer to the Modules tab in
2 Enter 1 for the SMU Slot Number.
3 Click the Save tab (right side of screen).
4 Right-click and rename as CV to IV. IMPORTANT: use this exact file name as
Testing the Switch Setup
1 Select the IV to CV Favorite.
2 Click the Recall button.
3 Click the green Play arrow or RUN > Single.
4 Click the Close button. You should hear clicking in the test chamber.
5 Select Tools > System > Configuration.
6 Select the UHC Expander / Fixture tab. The Input SMU field should be empty.
7 Select the CV to IV Favorite.
8 Click the Recall button.
9 Click the green Play arrow or RUN > Single.
10 Click the Close button. You should hear clicking in the test chamber.
11 Click Tools > System > Configuration.
12 Click the UHC Expander / Fixture tab. The Input SMU field should have the SMU
42
"Configure EasyEXPERT" on page 37. Typically, the HVSMU slot is Slot 7.)
shown, including the spaces between words and capitalization.
name.
Keysight PD1000A Power Device Measurement System Startup Guide
IV and CV Test Requirements