Keysight PD1000A Startup Manual page 77

Power device measurement system for advanced modeling
Hide thumbs Also See for PD1000A:
Table of Contents

Advertisement

Running the S-Parameter Characterization
2 The PD1000A Control Software system prompts you to calibrate the Network
Figure 38 Calibrate Network Analyzer Prompt
3 After initialization, etc. the test runs and the S-Parameter plots appear on
Keysight PD1000A Power Device Measurement System Startup Guide
Analyzer.
Calibrate the ENA with the appropriate PD1000A Test Fixture connected.
Use the Calibration Standards supplied with the test fixture. Refer to the
PD1000A Test Fixture User Guide for calibration details.
– If you have already calibrated the analyzer, click the Continue button.
– If you want to check the existing calibration or recalibrate for a different
DUT, click the Check Calibration button. This opens the calibration wizard on
the analyzer and adds four new buttons to the Calibration Check dialog.
These buttons are designed to help you with the calibration procedure.
-- Click the Open, Short, or Load button to display a correct calibration
pattern for the network analyzer during the calibration process.
-- Click Reset Calibration to reset all calibration settings for the analyzer
and fixture.
-- After calibrating, remove the calibration standards from the test fixture
and insert the DUT. Click Continue.
PD1000A software display (on the host PC). Refer to
On State and Off State S-Parameter Characterization
Figure 39 and Figure 39
.
75