IV and CV Device Characterization
Figure 24 Select the Tests You Want to Use
5 (Optional) Click the Select All Test check box to run all of the IV and CV tests.
6 Provide the device parameter information for each test. See "Setting Device
54
software help file for a full list.) Note that the screen graphic schematic
changes to illustrate the test that is running.
Parameters in the PD1000A Control Software" on page 57 below. For details
on these parameters, refer to the PD1000A Control Software Help File.
When setting the test parameters in the PD1000A test screens, do
not exceed device limits specified in the device's data sheet.
If you are not using the InTEST Thermal Platform to provide heat in
the B1506A Test Chamber, make certain the Temperature Control
Mode is set to Manual and set the temperature to 25 ° C in the
temperature field.
Gate Resistance. Specify the desired Gate Resistance for the device,
either 0 (default), 10, 100, 1000 . The Gate Resistor value must
be set in two places, here for the specific test and also in the
B1506A EasyEXPERT. See "Oscillations in the IV Plot," starting on
page 59. You must use the same values in both places.
Keysight PD1000A Power Device Measurement System Startup Guide
Running the IV and CV Tests