IV and CV Device Characterization

IV and CV General Test Setup

B1506A Setup

1 The B1506A SMU Measurement Resources depend on which option you
2 Connect the analyzer to its Test Chamber as described in the B1506A or
30
To avoid electrical shock and instrument damage, turn all
instruments off before connecting cables or inserting a device to
be tested.
purchased. Verify that the following SMUs are installed in the correct slots.
-- B1506A option H21 (20 A) has the following SMUs:
Slot 1: B1511B MPSMU (SMU1)
Slot 2: B1520A MFCSMU
Slot 3: B1514A MCSMU (SMU2)
Slot 4: B1514A MCSMU (SMU3)
Slot 5: B1514A MCSMU (SMU4)
Slot 6: Not Used
Slot 7: B1513C HVSMU (SMU6)
Slot 8: Not Used
-- B1506A options H51 (500 A) and H71 (1500 A) are identical with the
following SMUs:
Slot 1: B1511B MPSMU (SMU1)
Slot 2: B1520A MFCSMU
Slot 3: B1514A MCSMU (SMU2)
Slot 4: B1514A MCSMU (SMU3)
Slot 5: B1514A MCSMU (SMU4)
Slot 6: B1514A MCSMU (SMU5)
Slot 7: B1513C HVSMU (SMU6)
Slot 8: Not Used
The H51 and H71 options configure the UHCU (Ultra High Current
Unit) by using the test chamber to merge the currents of two
MCSMUs installed in the B1506A mainframe. The UHCU is not
available with option H21.
B1505A documentation. See Figure 6 below, which shows a B1506A.
a Connect the Digital I/O cable first.
b Connect the N1300 CMU cable; used on the MFCMU in the B1506A Slot 2.
c Connect the System cable in the order of the label numbers. On the
B1506A, for example, start at the bottom with the GNDU cable and work
Keysight PD1000A Power Device Measurement System Startup Guide
IV and CV Test Requirements