IV and CV Device Characterization
B150xA Specification Conditions
The measurement and output accuracy are specified under the conditions listed
below. Note: The SMU measurement and output accuracy are specified at the
output terminals in the test fixture except for capacitance measurement that is
specified at the output terminals of the MFCMU.
B150xA Operating Conditions
The B1506A or B1505A PDA must to be used under the conditions listed below.
28
– Temperature: 23 ±5 °C.
– Humidity: 20 to 70%, No condensation.
– Self-calibration after a 40 minute warm-up is required.
– Room temperature change less than ±1 °C after self-calibration execution.
(Note: This does not apply to the MFCMU).
– Measurement made within one hour after self-calibration execution. (Note:
This does not apply to the MFCMU).
– Calibration period: 1 year.
– SMU integration time setting: 10 PLC (1 nA to 1 A range, voltage range),
200 μs (20 A range) Averaging of high-speed ADC: 128 samples per 1 PLC.
– SMU filter: ON for MPSMU.
– The accuracy of the drain output current measurement specification is not
guaranteed until 20 seconds after a voltage change.
– Temperature: +5 to +40 °C
– Humidity: 20 to 70%, No condensation
– When used with InTEST Thermal plate:
-- Temperature: +5 to +30 °C
-- Humidity: 20% to 70%, No condensation
-- The B1505A requires the Keysight N1265A text fixture to work with the
InTEST platform.
Important: Always use the latest firmware in the analyzer. Firmware
updates are available at:
www.keysight.com/find/B1505A
Use EasyEXPERT Software version 6.1.1742.9630 (or later).
Software updates are available at:
www.keysight.com/find/EasyEXPERT.
www.keysight.com/find/B1506A
Keysight PD1000A Power Device Measurement System Startup Guide
or