Preparing the cable exchange
Replacing cables
1.
59 38 399 D3352
D3352.076.01.26.02 09.2016
6.23.2
Replacing cables (L3, L5, L6, L11, L12, and
L15)/Laying the cable/corrugated tube at the
rotation unit
CAUTION
The connectors and cables must be protected by inserting them in the fabric
tube A supplied with the cables.
1.
Remove the covers.
2.
Pull the connectors off board DX6.
A
2.
1.
For L12: Protect the tongue of connector RJ45 by wrapping it with insulat-
ing tape.
ATTENTION! Make sure that the contacts don't get sticky or stuck to-
gether!
2.
For L3: Remove the connector of the cable with tool W1.
3.
Remove the corrugated tube and the spiral spring B from the cable loom.
4.
Remove the defective cable and run the new cable up to the rotary ring in
the original position.
5.
Bunch the cables together again to form a loom.
6.
Fasten the defective cable to the loom and use it as a pull wire to pull the
loom through the fabric tube A.
7.
Pull the fabric tube over the connector and as far over the cable loom as
possible.
6.23 Replacing cables
W1
6 – 103